Introduction

The purpose of this post is to explain how to utilize medical imaging data in the development of a prosthetic implant. The two most common medical imaging technologies are CT and MRI. Both export a stack of 2D grey scale images over a 3D domain in the standard Digital Imaging and Communications in Medicine (DICOM) format. In this post I will go through the development of geometric (CAD) and mechanical (FEA) models based off anatomical imaging data. Through this workflow designs can be tuned for specific biometry based on realistic loading scenarios. As always all of the models used to develop this post are available at the end of the article.

Importing CT Scan

Inside Simpleware’s ScanIP, DICOM image data from a CT scan of a femur is imported. There are a variety of data sampling methods which are flexible enough for biomedical and even geological imaging. This post is going to focus on the major methodology steps and pass over details such as surface smoothing and filtering. Here, the raw DICOM data has been imported. The image stack has been partitioned based on the grey scale images and the solid bone geometry is shown below.

CT ScanIP Femur Translucent

Once an appropriate solid representation of the femur is created the model is then opened with Simpleware’s +CAD module.  An STL of the bone was exported so that the prosthetic implant can be appropriately located with respect to the femur within Catia.

 

Assembly of bone and implant

The STL of the femur is then imported into Catia using Digitized Shape Editor as a point cloud.

CT Caitia Femur

From here subsets of the femur cloud were created for the head and the medullary cavity (hollow center where the bone marrow is located).

CT Caitia Femur Sectioned

After these two subsets were created Quick Surface Reconstruction was used to fit a sphere and cylinder to them respectively. These will be used in the assembly to mate the implant to the scan data. The center point of the sphere and the axis of the cylinder were used to create a plane for mating as well.

CT Caitia Femur Fit

Here the implant has been located with respect to the femur. The following mates were created: spherical center points, planes created and the implant stem was mated to the distal end of the cylinder axis. A parametric implant model could be created to individually size the implant to the patient allowing for the axes to be mated. As you can see the stem of the implant is not perfectly aligned with the medullary cavity.

CT Caitia Femur Implant

Merging Femur and Implant

The located implant was then exported from the Catia assembly as a STEP file for import into Simpleware’s +CAD module.

CT ScanIP Femur Implant

The implant was then fit with voxels (3D pixels) to be compatible with ScanIP’s workflow. The femur head was then removed with a planar cut. The remaining femur and implant were merged together retaining the boundary. A mesh was applied to the model and was exported as an INP file (Abaqus’s input file format).

CT ScanIP Femur Implant Sectioned

Finite Element Model

Once in Abaqus a simple FEA model was created. The distal end of the femur was fixed and a pressure load was applied to the superior portion of the prosthetic head, simulating a standing load.

CT Abaqus Femur Implant Mesh

Simulation Results

A critical design feature of implants is the need for load sharing between the implant and remaining bone. If the implant is too stiff there is load shielding and the bone’s density will reduce. If the implant is too compliant there will be significant contact stresses. As with most critical design features a balance must be achieved and numerical simulation aides the designer in making informed engineering decisions. Several strain plots of the implant and bone are shown below.

CT Abaqus Femur Implant Strain2CT Abaqus Femur Implant Strain1

Conclusion

The use of ScanIP, Catia and Abaqus provide the robust tools needed for orthopedic design. ScanIP enables use of the 3D image data that CT and MRIs produce. With this data implants can be appropriately sized and located with respect to the patient data in Catia. Simulating the mechanics of the system in Abaqus with FEA then drives design changes such as stem stiffness and cross sectional shape.

Realistically simulating the geometry and physics of the hip prosthesis system design studies can be made in silico instead of in vivo. This avoids costly clinical trials with respect to time and money for exploitative design studies. The end goal is to simulate with enough accuracy and confidence that the clinical trial is merely a formality to make sure nothing has been overlooked. This technology enables the ability to provide customized implants based on biometric data.

I hope you found this informative please feel free to comment, like, subscribe, contact us or whatever else can be done today! Thank you.